12 Generator Calibration<br />

ACCESS CALIBRATION SOFTWARE<br />

1. Press The SETUP OPTIONS key on the Workstation keyboard. The Setup<br />

Options screen will appear.<br />

2. Insert the OEC service access disk into the floppy disk drive on the rear of the<br />

Workstation.<br />

3. Select Access Level 2 and press the ENTER key.<br />

4. Select the Access Mainframe Menu and press the ENTER key.<br />

The Mainframe Menu will appear on the left monitor. The Control Panel Display will<br />

display “MAINFRAME MENU.”<br />

A - Run Applications (Take X-rays)<br />

B - Run Calibration With Parameters<br />

C - Run Calibration<br />

D - System Configuration<br />

E - mA Limit File Editor<br />

F - Examine Event File<br />

G - Run Status<br />

H - Collimator Calibration<br />

I - Language Options<br />

J - ABS Index File Editor<br />

- Return to DOS<br />

Enter Command Letter :<br />

MAINFRAME MENU<br />

5. From the Mainframe Menu select: C - Run Calibration. The Control Panel Display<br />

will display “CALIBRATION”.<br />

6. Enter your initials when prompted: "Enter Initials Please".<br />

7. Answer the following question when it appears on the monitor by entering Y and<br />

pressing the ENTER key:<br />

Is your Dynalyzer (if any) wired internally for automatic reset?<br />

Enter (Y/N): Y