Peripherals 7<br />

VCR CONFIGURATION<br />

Verify the VCR is configured properly in software by performing the following steps:<br />

1. Power on the Workstation and allow it to boot <strong>com</strong>pletely.<br />

2. Insert the Diagnostic Diskette in the rear of the Workstation.<br />

3. Press the SETUP OPTIONS key on the Workstation Keyboard as shown in<br />

Figure 5.<br />

PATIENT DISK IMAGE<br />

HELP (ANNOTATE) VIEW DIR<br />

X-RAY SPECIAL TEST<br />

SETUP<br />

MARKERS SUMMARY APPS PATTERN CUSTOMIZE OPTIONS<br />

Figure 5 - “SETUP OPTIONS” Key on Workstation Keyboard<br />

4. Select Access Level 2 and then the System Configuration option. A menu as<br />

shown in Figure 6 appears:<br />

Set configuration<br />

Line Frequency (Hz) 60<br />

System Model VAS<br />

Cine Disk ALTA4<br />

Language ENGLISH<br />

VCR1 Sony<br />

VCR2 Sony<br />

Digital Print YES<br />

Serial Number: xx-xxxx<br />

Hospital Name:<br />

Figure 6 - System Configuration Menu