Image System 31<br />

CCD Thermo-Electric Cooling Circuit (TEC)<br />

The CCD sensor is cooled by a thermo-electric cooling (TEC) and heat sinking<br />

apparatus. The cooling module attempts to maintain the CCD temperature at the same<br />

temperature as the image intensifier housing. Poor image quality may indicate that the<br />

TEC circuit is not working properly.<br />

The heat sinking apparatus is turned off and on periodically by firmware located on the<br />

Image Function PCB. The ON and OFF cycle times depend on how long the processor<br />

has been running.<br />

A +5 VDC is applied to R4 on the Image Function PCB. Current flows through R4 to the<br />

thermo-electric cooler, back into the Image Function PCB and through transistor Q3 to<br />

DC <strong>com</strong>mon ground. Micro-controller U16 pin 19 controls the base of transistor Q3. A<br />

high on pin 19 turns the cooler ON and a low turns the cooler OFF. These times have<br />

been determined experimentally and are stored in a look-up table (LUT) on the Image<br />

Function PCB. To determine proper operation of the TEC circuit, measure TP3 on the<br />

Image Function PCB for a drop to +2.4 VDC ± 0.75 V. The E9 jumpers are not used at<br />

this time.<br />

SYNCHRONIZATION STANDARDS<br />

A 60 Hz and 50 Hz model CCD Camera is available.<br />

Domestic Models - RS-170A Video Waveform Standard, 60 Hz.<br />

International Models - CCIR or PAL <strong>com</strong>patible with 576 lines centered in the<br />

unblanked area of the CCIR standard, 50 Hz.