C-Arm Software 9<br />

CAUTION: Use this option to ignore the stator sense circuit.<br />

E. (NOCAL) Run machine uncalibrated.<br />

Do not use this option.<br />

F. (HARDWARE) Ignore A/D checks during boot-up.<br />

CAUTION: Use this option to ignore A/D failures.<br />

G. (CALCOLL) Calibrate collimator.<br />

Use this option to calibrate the collimator. See the Image Calibration section for details<br />

on this procedure.<br />

H. Change Contactor Engage Voltage (enter number):<br />

CAUTION: Do not use this option under any circumstances.<br />

I. Delete last parameter.<br />

Use this option to delete the last parameter that appears on the same line as option K.<br />

(RUN).<br />

J. (HEAT) Display Heat Info.<br />

Do not use this option.<br />

K. (RUN) MAIN<br />

Use this option to start the generator application software when the desired parameters<br />

have been selected. This parameters appear on this same line.