10 C-Arm Software<br />

C - Run Calibration<br />

Refer to Generator Calibration regarding this option.<br />

D - System Configuration<br />

When this option is selected, the Control Panel displays “RUNNING OPTUSER” and the<br />

following menu appears.<br />

Creating personality file for <strong>9600</strong> Mainframe<br />

A. Is boost available [Y/N]? Y<br />

B. Is X-ray Inhibit line active low [Y/N]? Y<br />

C. Fluoro Beeper ON [Y/N]? Y<br />

D. Video Level set at: 450<br />

E. Percent of maximum available boost: 90<br />

F. Fluoro timer set for [5-10] minutes 5<br />

G. Boosted Pulse Available [Y/N]? Y<br />

H. Film shots allowed on boost button only [Y/N]? N<br />

I. Pulse Video Level set at: 315<br />

J. Video level dead band: 16<br />

K. Boost time: 30<br />

L. Default ABS Tables: DOMESTIC 1<br />

Exit Configuration and Update personality file<br />

Enter desired choice:<br />

NOTE: Options B, D, & J should appear as shown above. Other options may differ.<br />

NOTE: [Y/N] choices toggle when option letter (A, B, etc.) is pressed.<br />

E - mA Limit File Editor<br />

Refer to Generator Calibration section regarding this option.