Image System Calibration 27<br />

NOTE: If the Bore Sight Tool is not centered in the image, proceed to step 6.<br />

Otherwise, if the Bore Sight Tool is centered in the image, proceed to step<br />

8.<br />

6. Loosen the three screws securing the secondary collimator to the nylon gear on the<br />

variable collimator as shown in Figure 27.<br />

Figure 27 - Secondary Collimator Mounting Screws<br />

7. While making an exposure adjust the secondary collimator position until the bore<br />

sight is centered on the pattern image as shown in Figure 26.<br />

8. Tighten the screws on the secondary collimator.<br />

9. Remove the bore sight tool to perform the next procedure.