Image System Calibration 31<br />

9. Select MAG1 field size and while making an exposure, press and release the<br />

collimator iris increment/decrement key until the exposed image is within the “B”<br />

ring indicated on the Beam Alignment Tool pattern as shown in Figure 30.<br />

Figure 30 - "B" Ring of Beam Alignment Tool<br />

A. The flat surfaces should be within the ring. After setting the iris position, remove<br />

any iris “gear play” by manually turning the iris gear in the open direction (without<br />

disturbing the iris servo loop).<br />

B. If the servo loop is disturbed, select another field size, return to MAG 1 and<br />

repeat this step. With the gear play removed, verify the iris remains within the<br />

“B” ring. If not, repeat this step until the iris (with gear play removed) is within the<br />

“B” ring.<br />

NOTE: If any inward bow is observed along the “flat edges” of the collimator, the<br />

measurement must be made across the flats as though the flats are<br />

straight.