Workstation Control 15<br />

COMMUNICATIONS PCB<br />

Assy. #00-872127 - Sch. #00-872125<br />

The Communications PCB provides serial and parallel <strong>com</strong>munications capabilities for<br />

the Motherboard with external devices, such as a printer. Internal system<br />

<strong>com</strong>munications with the Control Panel Processor PCB and the C-Arm are also handled<br />

by the PCB. The circuit board plugs directly into the AT Motherboard (see Figure 4).<br />

SERIAL COMMUNICATIONS<br />

The Communications PCB handles all of the serial <strong>com</strong>munications for the AT<br />

Motherboard to the Workstation Control Panel Processor PCB and the C-Arm.<br />

AT Motherboard to Control Panel Processor<br />

Signals from the 386 AT Motherboard are received by DUART U27 on the AT<br />

Communications PCB via the data bus. U27 converts this information to RS232 serial<br />

data. The serial data then passes through the Aux Interface PCB, through RS232<br />

transceiver U7 on the Control Panel Processor and to microcontroller U5. Microcontroller<br />

U5 on the Control Panel Processor PCB <strong>com</strong>municates over the same path to the AT<br />

Motherboard<br />

UART U26 on the AT Motherboard handles discrete signals for the room interface relays<br />

on the Aux Interface PCB when they are used.<br />

AT Motherboard to C-Arm<br />

Communication between the C-Arm and Workstation is established during the boot<br />

process. All 20 arrows will be displayed on the Control Panel Display, as shown below,<br />

during this part of the boot process.<br />

During normal operation, the Technique Processor and 386 AT Motherboard send<br />

information such as:<br />

• Shot log file information from the C-Arm to the Workstation<br />

• Commands such as L>R, SAVE and CART MODE from the Control Panel to the<br />

Workstation<br />

• “Enable Boost” <strong>com</strong>mand to the C-Arm when (CHOLE)/DIGITAL SPOT is selected on<br />

the Workstation<br />

The TXD and RTS signals from Port B of the serial <strong>com</strong>munications controller U39 on<br />

the Technique Processor PCB are sent to the 386 AT Motherboard on the Workstation.<br />

They pass through RS232 driver/receiver U52 on the Technique Processor PCB and<br />

then pass through the Mainframe Motherboard, the Interconnect cable, and enter the