Image System Calibration 25<br />

COLLIMATOR MECHANICAL CENTER<br />

WARNING... Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. Press ESC four times to remove the Monitor/Camera Alignment Pattern. Then<br />

select the MAG 1 field size.<br />

2. While making an AUTO FLUORO exposure, close the collimator until it is<br />

approximately the size of the C-ring on the Beam Alignment Tool pattern as shown<br />

in Figure 24.<br />

Figure 24 - MAG 1 Collimator Sizing<br />

3. Adjust the X and Y screws on the Beam Alignment Tool to center the collimator<br />

around the C-ring. If necessary, reposition the entire collimator assembly to get the<br />

image on the CRT.