Image System Calibration 29<br />

VARIABLE COLLIMATOR CENTERING<br />

WARNING... Steps within this procedure produce X-rays. Use appropriate precautions.<br />

1. Select the NORMAL field size<br />

2. Reinstall the Bore Sight Tool.<br />

3. Take an exposure to determine whether the Variable Collimator requires<br />

adjustment. The bore sight should appear centered on the image.<br />

4. If the Variable Collimator does not require adjustment continue to the next section<br />

"Collimator Iris Calibration" Otherwise, perform the following:<br />

A. Loosen the 4 collimator mounting screws, shown previously in Figure 19, by<br />

turning them approximately 1/8 of a turn CCW.<br />

B. While making an exposure, adjust the collimator positioning plate with the X and<br />

Y adjustment screws and center the X-ray beam in the Beam Alignment Tool<br />

pattern as shown in Figure 28.<br />

Figure 28 - Positioning the Variable Collimator<br />

C. Tighten the collimator anchor screws (two with large flat washers) and then verify<br />

alignment with another exposure.<br />

D. Tighten the remaining collimator screws and remove the Bore Sight Tool.<br />

5. Reinstall the secondary filter on the collimator using the four washers and nuts.