12.01.2013 Views

9600 Service Manual - gmecorp-usa.com

9600 Service Manual - gmecorp-usa.com

9600 Service Manual - gmecorp-usa.com

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

mA Generation 13<br />

MA TROUBLESHOOTING<br />

MA CONTROL<br />

With 80 kV @ 5 mA selected in the manual fluoro mode and no X-ray switch pressed,<br />

you should measure approximately -1.7 VDC on TP8 of the X-ray Regulator PCB. This<br />

voltage should go more negative as the mA setting is increased and less negative as the<br />

mA setting is decreased. No X-rays are necessary to perform this check.<br />

The ratio for TP8 is approximately -350 mV/mA in the fluoro mode and approximately -<br />

70 mV/mA in the film mode.<br />

MA DRIVE<br />

In the manual fluoro mode with no X-ray switch pressed, the FILDRVA and FILDRVB<br />

input signals to PAL U10 (pins 5 & 6) on the X-ray Regulator PCB should be out of<br />

phase with a frequency of 2.5 kHz. The duty cycle should decrease as less mA is<br />

selected and increase as more mA is selected. The FILDRA and FILDRB output signals<br />

from PAL U10 pins 12 and 13 should do the same. No X-rays are necessary to perform<br />

this check.<br />

FILAMENT REGULATOR<br />

In the <strong>Manual</strong> or Auto Fluoro mode with no X-ray switch pressed, you should measure<br />

approximately +4 VDC on U27-6 (FIL B+ CTRL) of the X-ray Regulator PCB and<br />

approximately +160 VDC out of the filament regulator on R8 (or Q4 collector) of the<br />

Generator Driver PCB. The FIL B+ SNS voltage on U27-1 of the X-ray Regulator PCB<br />

should also measure approximately +4 VDC. All three of these voltages should vary<br />

during an exposure.<br />

The ratio for FIL CNTRL is approximately 1:40, for every +1 VDC on TP10 of the Analog<br />

Interface there should be approximately +40 VDC out of the filament regulator.<br />

MA SENSE<br />

During an 80 kV @ 5 mA <strong>Manual</strong> Fluoro exposure, you should see mA voltage pulses<br />

on TP7 of the X-ray Regulator PCB with an amplitude of approximately +950 to<br />

+1200mV (using an oscilloscope). However, the average DC voltage <strong>com</strong>ponent of<br />

these voltage pulses on TP7 should be approximately +125 mVDC (measured with a<br />

DC Volt meter). The voltage on TP6 should be approximately +125 mVDC and TP11<br />

(MA Measured) should be approximately +625 mVDC. All of these voltages should<br />

increase as actual tube current (mA) increases and decrease as actual tube current<br />

(mA) decreases.<br />

The ratio for TP6 and TP7 is +40 mA/VDC in both Fluoro and Film modes. The ratio for<br />

TP11 is +8 mA/VDC Fluoro mode and +40 mA/VDC Film mode.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!