12.01.2013 Views

9600 Service Manual - gmecorp-usa.com

9600 Service Manual - gmecorp-usa.com

9600 Service Manual - gmecorp-usa.com

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

10 Image System Calibration<br />

MONITOR/CAMERA ALIGNMENT PATTERN<br />

1. With the system booted, press the SETUP OPTIONS key on the Workstation<br />

keyboard. The Setup Options menu will appear.<br />

2. Insert the OEC Boot and Diagnostic disk into the disk drive.<br />

3. Select Access Level 2 (<strong>Service</strong>) and press the ENTER key.<br />

4. The <strong>Service</strong> Options menu will appear.<br />

5. Enter the Video Calibration sub-menu, select Monitor/Camera Alignment and<br />

press the ENTER key.<br />

6. The Monitor/Camera Alignment Pattern, shown in Figure 8, is displayed on both<br />

monitors.<br />

Figure 8 - Monitor/Camera Alignment Pattern<br />

7. This pattern will be used at various times for the remainder of this procedure.<br />

8. If the monitors do not show a linear and round image as shown in Figure 8, it may<br />

be necessary to adjust the size, centering, and linearity. Proceed to the next portion<br />

of this procedure, "Confirm Monitor Size & Linearity".