May 6, 2005 <strong>iCys</strong> Cytometric Analysis Software Application Overview<br />

Width—The distance between the first and second background contour. Set<br />

the width such that the outer contour line includes the dimmest pixels adjacent<br />

to an event.<br />

Use pixels between—When determining the background, you usually want to use<br />

those pixels that represent the mean pixel value, such as 30-70%. If, however, the<br />

background is very bright, you might want use a lower threshold, such as 20-50%.<br />

Color—Select the color of the background contour lines. If color is set to None,<br />

no visible contours will be drawn.<br />

Corrected Channels—Allows you to select the individual channels that will use<br />

the background correction.<br />

Peripheral Tab<br />

Selecting the Peripheral tab opens the following menu.<br />

Peripheral contour—Defines two boundary lines that are used to determine the<br />

periphery of an event.<br />

Distance—Specifies the distance between the integration contour and the first<br />

peripheral contour. Set the first peripheral contour such that it picks up the<br />

brighter peripheral pixels.<br />

Width—Specifies the distance between the first and second peripheral<br />

contour. Set the width such that the outer contour line includes the dimmer<br />

pixels that are still within the periphery of an event.<br />

Details on the Contour, Background, and Peripheral tabs are found in “Integration<br />

Contours” on page 4-7, “Background Contours” on page 4-8, and “Peripheral<br />

Contours” on page 4-9.<br />

Phantom<br />

Phantom contours are events generated by <strong>iCys</strong> from circular contours applied<br />

over the laser-scanned image.<br />

Creating or clicking on a Phantom module opens the following menu.<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> 3-57