Installing <strong>iCys</strong> Research Imaging Cytometer General Installation Information<br />

General Installation Information<br />

Installing <strong>iCys</strong> Research Imaging Cytometer requires the following:<br />

❑ Sufficient space to accommodate the physical <strong>iCys</strong> equipment<br />

❑ Proper electrical power configuration and receptacles<br />

❑ Acceptable environmental conditions<br />

Note: <strong>iCys</strong> needs a vibration-free bench top. (Shock mounting is not required.)<br />

Unpacking the <strong>iCys</strong> System<br />

The <strong>iCys</strong> should be unpacked and installed only by a representative from CompuCyte<br />

or an authorized agent. If the system is to be stored temporarily before installation, it<br />

should be in a secure, heated, and dry area.<br />

Upon receipt of the <strong>iCys</strong>, examine the shipping materials for exterior damage or<br />

evidence of mishandling. If they appear damaged, notify the carrier and CompuCyte<br />

or an authorized agent and request an inspection from the carrier.<br />

Following initial installation, a representative from CompuCyte Corporation or an<br />

authorized agent will instruct designated personnel on the basic operation and care of<br />

the <strong>iCys</strong>.<br />

Physical Space Requirements<br />

The following table specifies the <strong>iCys</strong>’s space requirements:<br />

Table A-1: <strong>iCys</strong> Physical Space Requirements<br />

Analysis Unit ❑ Width 44 in. (112 cm)<br />

❑ Depth 32 in. (82 cm)<br />

❑ Height 31 in. (79 cm)<br />

To allow for servicing, the space above the table surface<br />

should be at least 44 inches or (112 cm)<br />

❑ Weight 275 lbs. (125 kg.)<br />

Computer ❑ Width 8 in. (21 cm)<br />

❑ Depth 19 in. (49 cm)<br />

❑ Height 17 in. (44 cm)<br />

Monitor ❑ 20” Flat panel display monitor<br />

20” (51cm) wide, 12” (31cm) depth, 20” (51cm) height<br />

❑ Dual monitors are optional<br />

Autofocus<br />

monitor (if<br />

installed)<br />

❑ Width 9 in. (23 cm)<br />

❑ Depth 10 in. (26 cm)<br />

❑ Height 9 in. (23 cm)<br />

A-2 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B