Administrator Tasks Defining Custom Carriers<br />

For example:<br />

Carrier Side View<br />

Bottom<br />

Thickness<br />

Base to<br />

Well Height<br />

Outside edge of carrier<br />

{<br />

If you do not know the measurements for Base to Well Height or Bottom<br />

Thickness, use the method described in “Determining the Microplate Z<br />

Parameters” on page C-14. When you have entered these measurements,<br />

return to Focus Methods, below.<br />

✧ Retraction Offset—Only used with some Freeform Carriers. For more<br />

information, see “Defining a Freeform Carrier” on page C-16.<br />

❍ Focus Methods—Specify the preferred (or required) focus method.<br />

✧ Bottom then Top—For thicker plastic-bottomed plates (greater than<br />

400µ) especially; <strong>iCys</strong> will focus first on the plate bottom, then translate<br />

up by the bottom thickness to the plate top and refocus.<br />

✧ Bottom Only—For glass-bottomed or thinner plastic-bottomed (less than<br />

400µ) plates, <strong>iCys</strong> focuses directly on the bottom then moves up by the<br />

bottom thickness plus other offsets to bring the sample into focus.<br />

✧ Default—To allow your users to select a focus method (as discussed in<br />

“Focus and Performance” on page 3-33), select both focus methods and<br />

then select the method that you want to use as a default setting.<br />

Note: If the Bottom Thickness is less than 400µ, select Bottom Only as the only<br />

available focus method. <strong>iCys</strong> will set the default focus method to match.<br />

❍ Test Autofocus—To test whether you have the values for your carrier set<br />

correctly, you can run Test Autofocus. To test under working conditions,<br />

CompuCyte suggests that you fill the test wells with water or another buffered<br />

solution (such as phosphate-buffered saline [PBS].)<br />

Test Autofocus tries to focus the objective correctly on the carrier and<br />

produces a result number, as follows:<br />

✧ If the Test Autofocus Result is greater than 800, the parameters are<br />

correctly set.<br />

✧ If Autofocus did not work, an error message will display. Check the<br />

General tab of the Instrument Setup window (and the Focus parameter in<br />

the task bar) to ensure that Autofocus is selected. You may also need to<br />

check the values of the X,Y, and Z parameters for errors.<br />

✧ If the Test Autofocus Result is between 0 and 800, try to refine the values<br />

for the carrier and retest.<br />

7. When you are happy with the results, click Save to save your changes. Close the<br />

Edit Carrier Type window.<br />