May 6, 2005 <strong>iCys</strong> Cytometric Analysis Software Application Overview<br />

❍ In the Scan Cycle Time field, enter the minimum number of seconds between<br />

subsequent scan start times. If the scan finishes before the cycle time has<br />

elapsed, <strong>iCys</strong> will wait until the next cycle to start scanning again. This allows<br />

you to, for example, record snapshots at specific intervals of time.<br />

Focus and Performance<br />

Clicking the Focus and Performance icon ( ) opens the following menu.<br />

The Focus and Performance menu functions are described below.<br />

Focus<br />

Allows you to select whether you want to focus the microscope manually or allow<br />

<strong>iCys</strong> to AutoFocus for each scan area:<br />

❍ AutoFocus—If your system uses the optional AutoFocus feature, a<br />

✧ Focus Method—Specify the preferred focus method.<br />

✧ Bottom then Top—For thicker plastic-bottomed plates (greater than<br />

400µ) especially; <strong>iCys</strong><strong>iCys</strong> will focus first on the plate bottom, then<br />

translate up by the bottom thickness to the plate top and refocus.<br />

✧ Bottom Only—For glass-bottomed or thinner plastic-bottomed (less<br />

than 400µ) plates, <strong>iCys</strong> focuses directly on the bottom then moves up<br />

by the bottom thickness plus other offsets to bring the sample into<br />

focus.<br />

If the carrier type supports only one focus method, the supported method<br />

will be selected and the unsupported method will be grayed out and<br />

unavailable.<br />