May 6, 2005 Administrator Tasks<br />

❍ A well at or near the center top row<br />

❍ A well at or near the center bottom row<br />

❍ A well at or near the center left column<br />

❍ A well at or near the center right column<br />

For example:<br />

5. Determine which of these 9 positions yields the lowest value, and enter that value<br />

as the Base To Well Height.<br />

Determining the Bottom Thickness<br />

➣ To determine the bottom thickness:<br />

1. Move the stage so that the objective focuses on the well with the lowest Base to<br />

Well Height (which you determined in “Finding the Base to Well Height” on<br />

page C-14), and click on Move To from the Base to Well Height field.<br />

2. Make sure the well you use for focusing is empty.<br />

x<br />

x<br />

x<br />

3. Slowly rotate the fine focus knob counter-clockwise until the second line or lines<br />

appear to be finely focused on the autofocus monitor. For thicker-bottomed plates,<br />

this may take 5 or more knob rotations.Click Set from Z for the Bottom<br />

Thickness field of the Z tab.<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> C-15<br />

x<br />

x<br />

x<br />

x<br />