<strong>iCys</strong> Cytometric Analysis Software Application Overview <strong>iCys</strong> Task Bar<br />

number of X Steps to generate a square scan field. The scan field and pixel<br />

size are determined by the dimensions you select and are displayed in the<br />

Scan field dimension window below.<br />

❍ Scan field dimension—Displays the pixel size and size of the scan field<br />

based on the parameters you have selected.<br />

❍ Stop on count—Checking this box will stop the scan when the number of<br />

events in the designated Region reaches the designated count.<br />

Note that when the count has been reached, the system might scan one extra<br />

field since events are analyzed while scanning the next field. Events from the<br />

extra field will be ignored when stop on count has been reached.<br />

❍ Advanced Tab—Clicking the Advanced tab opens a menu in which you can<br />

specify whether or not to apply pixel correction, the Y interval in microns,<br />

whether or not to save field images, and, if so, in what format. You can also<br />

choose to save mosaic images for a Test Scan, a scan and save, or both.<br />

❍ Scan Mode—Describes current scan (Virtual, Live, Bypass)<br />

❍ Correct Pixels—Checking this option enables scanline correction to account<br />

for variations in collection efficiency along the scanline. This is performed on<br />

each of the images as they are acquired, and is shown in the figure below.<br />

3-50 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B