Using the iNovator Toolkit Multi-scale Scanning<br />

Image Hierarchy<br />

The first scale Mosaic Scan image is displayed in the Well Image window. As the<br />

second scale Field Scan progresses a mosaic image of each Dynamic Scan Area is<br />

displayed in the Region Image window. Each scan field image is displayed in the Field<br />

Image window.<br />

From the Well Image, clicking on a particular Scan Area Event will display that event<br />

in the Region image window. Similarly, clicking on a particular scan field location in<br />

the Region Image will display that scan field in the Field Image window.<br />

Analysis Modes for Multi-scale Scans<br />

New Runs<br />

When setting up a new run or re-running an existing run, the user has several<br />

operating modes to choose from. What is meant by operating modes is this: Will the<br />

systems be acquiring new data by scanning a sample or will the system be reanalyzing<br />

previously acquired images by performing a virtual scan of those images? With multiscale<br />

scanning, certain combinations of these two modes are possible. All of these<br />

options are described below.<br />

Live-Live<br />

With a new run, data must be acquired with a live scan. If a fully automated run is<br />

chosen then data will be acquired for both scales by live scanning of the sample. To set<br />

up a fully automated 2 scale scan, make sure the bridge between the two Scan modules<br />

is NOT set to pause by clicking on the box on the bridge until it appears blank, as<br />

shown in the figure below.<br />

Test scans should be made to set the hardware and software protocol parameters for<br />

both scale scans. To set up the protocol for both scales create an abbreviated scout<br />

scan area, turn on the pause between scales and select the Test Scan button or the<br />

Next button (single field). During the scout scanning make the desired changes to the<br />

PMT settings. (NOTE: This may take several iterations of starting and stopping the<br />

test scan.) When the PMT settings are complete, stop the scan and then acquire<br />

another Scout scan using the correct PMT settings. During the Pause, use the View<br />

Image tool as described in “Using the Protocol Menu View Image Feature with<br />

iNovator” of this appendix. Once you are satisfied with the contouring protocol, select<br />

or create Dynamic Scan Areas in the Well Image window to use for setting up the 2nd<br />

scale hardware and contouring protocol. Delete any unnecessary dynamic scan areas.<br />

(NOTE: You may want to increase the size of the dynamic scan areas to be large<br />

enough to complete the 2nd scale setup.) Once you have selected the dynamic scan<br />

areas, press the Paused button to continue the test scan in the 2nd scale. During test<br />

scanning of the 2nd scale make the necessary hardware and contouring protocol<br />

changes. (Note: It may take several iterations of test scanning to accomplish the<br />

setup).<br />

B-34 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B