Administrator Tasks Defining Custom Carriers<br />

Well section of the XY tab to set the location. The X and Y margin parameters are<br />

automatically filled in by <strong>iCys</strong>.<br />

Note: Always turn the oscilloscope off and wait for the Current X and Current Y<br />

values (at the bottom of the XY tab) to update before you click Set from<br />

X,Y.<br />

The update may take a few moments, but will ensure that the<br />

measurement is correct.<br />

Setting the Interval using <strong>iCys</strong><br />

➣ To set the interval, as described in Step 5, above:<br />

1. Turn on the oscilloscope trace, and open the Argon Laser Shutter, then select B2<br />

from the Well pull-down list in the Other Wells section of the XY tab.<br />

2. View the beam, using a piece of thin white paper placed over well B2, and adjust<br />

the stage so the beam is centered in the well.<br />

3. Turn off the oscilloscope and click the Set from X,Y button in the Other Wells<br />

section. The Interval parameter is automatically filled in by <strong>iCys</strong>.<br />

4. To make sure that the interval is correct, select the last well H12 for a 96-well<br />

plate) in your microplate from the Other Wells Well pull-down list and click<br />

Move To to drive the stage to that well. Check that the beam is centered on the<br />

last well, again using thin white paper. If it is not centered, adjust the interval (but<br />

do not use the last well) and check again using the last well.<br />

Determining the Microplate Z Parameters<br />

Finding the Base to Well Height<br />

➣ To determine the Base to Well Height:<br />

1. Click on the Z tab in the Edit Carrier window.<br />

2. With the stage at well A1, the same objective in place (as displayed in the task<br />

bar), and the Oscilloscope window open and on, and the Argon Laser Shutter<br />

open, use the rocker switch on the Remote Control unit to drive the objective<br />

upwards slowly until you observe a reasonably well focused line or lines on the<br />

autofocus monitor. Make sure that you are focusing on the first line or pair of lines<br />

to be imaged on the display.<br />

3. Use the Fine Focus knob on the Remote Control unit to create a well focused<br />

image.<br />

4. Click Set from Z for the Base To Well Height field in the Z tab. Mark down this<br />

value, and then take the same measurement at a number of other points on the<br />

plate, such as:<br />

❍ The other three corner wells<br />

❍ A well near the center of the plate<br />

C-14 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B