Using the iNovator Toolkit Multi-scale Scanning<br />

Multi-scale Scanning<br />

Introduction<br />

Unique Features<br />

Multi-scale scanning involves defining a protocol for two sequential scans referred to<br />

as scales (because they are usually done at different resolutions). Typically, the first<br />

scale is done using the Mosaic Scan module at low resolution. This is referred to as a<br />

scout scan because it is scouting for relatively large structures such as tissue<br />

microarray elements, tissue sections, structures within tissue sections or cell colonies.<br />

Events found in this scout scan may be designated Scan Area Events. These scan area<br />

events become the defined scan areas for the second scale, typically higher resolution<br />

scan. Such a scan is defined in Tutorial 4, the protocol of which is shown below.<br />

Different Detector Module Settings<br />

The two scales' Detector module settings are independent of one another. Data<br />

channels, virtual channels, detector gain and offset, laser selection and (if applicable)<br />

laser power may be independently selected and set for each scale. This allows the user<br />

to set up the system for the very different scan requirements of scout vs. high<br />

resolution scans.<br />

Different Scan Module Settings<br />

The scan module settings are at the heart of defining the operation of each of the two<br />

scales. As was mentioned, the first scale is a Mosaic Scan, typically at low resolution<br />

B-30 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B