May 6, 2005 <strong>iCys</strong> Research Imaging Cytometer Hardware Reference<br />

Functional Specifications<br />

The following specifications apply to the <strong>iCys</strong> system.<br />

Microscope<br />

Motorized stage<br />

Table 2-2 describes the specifications for the microscope used with <strong>iCys</strong>.<br />

Table 2-2: <strong>iCys</strong> Microscope Specifications:<br />

Olympus IX-71 microscope<br />

base<br />

Objectives ❑ 10x magnification — 0.3 N.A.<br />

❑ 20x magnification — 0.5 N.A.<br />

❑ 40x magnification — 0.75 N.A<br />

The range and resolution of the motorized stage is:<br />

❑ Travel range:<br />

❍ X axis = 119.3mm<br />

❍ Y axis 116.3mm<br />

❑ Resolution—0.5 microns<br />

Crosshair alignment<br />

Eye Viewing WH10x eyepieces, F/22<br />

Video Viewing Sony CCD Video Camera, with EPIX frame<br />

grabber card<br />

Brightfield Illumination<br />

Assembly<br />

Optional EpiFluorescence<br />

Illumination Assembly<br />

❑ 100W precentered Halogen bulb<br />

❑ Brightfield power source, TH4-100<br />

❑ NA 0.55, 27mm Working Distance<br />

condenser with 5 position manual turret<br />

❑ Illumination Column allows 30 degree<br />

tilting<br />

❑ 100W Mercury Burner<br />

❑ Mercury Burner power source<br />

❑ Filter Slider with three filter pockets<br />

❑ WIG, WIBA, and NV filter cubes installed<br />

in Fluorescence Mirror Unit Cassette<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> 2-31