May 6, 2005 Administrator Tasks<br />

The Z tab allows you to specify, or determine, the appropriate start AutoFocus<br />

positions for the carrier. The parameters are:<br />

❍ Setup—Allows you to move the stage to a well to calibrate or to the Clear<br />

position.<br />

✧ Room—Allows you to specify which room you want to use to calibrate<br />

the Z parameter. Select the room in the Room pull-down and click Move<br />

Stage To. The stage moves to the specified room.<br />

✧ Move Objective to Clear—Click to move the objective to the Clear<br />

position.<br />

❍ Configurable Positions—Specifies the required values for the carrier height<br />

(the Z parameter):<br />

✧ Base to Well Height—Specifies the distance from the base of the carrier<br />

to the bottom of the room.<br />

✧ Bottom Thickness—Specifies the thickness of the glass or plastic<br />

through which the microscope needs to focus. If you use #1 coverglass<br />

with this carrier, the value is 0.12.<br />

For example:<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> C-19