Administrator Tasks Defining Custom Carriers<br />

Carrier Side View<br />

Bottom<br />

Thickness<br />

Base to<br />

Well Height<br />

Outside edge of carrier<br />

{<br />

Rooms<br />

If you do not know the measurements for Base to Well Height or Bottom<br />

Thickness, use the method described in “Determining the Freeform Z<br />

Parameters” on page C-22. When you have entered these measurements,<br />

return to Focus Methods, below.<br />

✧ Retraction Offset—The Retraction Offset is only needed if you have a<br />

carrier that may be bumped by the microscope when moving from room<br />

to room (the only standard carrier for which this is an issue is the 3 Room<br />

Petri Dish). If a Retraction Offset is required, but you do not know it, you<br />

can find it as follows:<br />

– Click on Instrument ⇒ Home Objective. Manually move the stage to a<br />

position between rooms using the touchpad or Remote Control unit<br />

knobs.<br />

– As described in “Finding the Base to Well Height” on page C-22, use<br />

the Remote Control unit knobs and the autofocus monitor to focus on the<br />

top of the carrier.<br />

– When you have a clear focus, click Set from Z for the Retraction Offset<br />

value.<br />

❍ Focus Methods—Specify the preferred (or required) focus method.<br />

✧ Bottom then Top—For thicker plastic-bottomed plates (greater than<br />

400µ) especially; <strong>iCys</strong> will focus first on the plate bottom, then translate<br />

up by the bottom thickness to the plate top and refocus.<br />

✧ Bottom Only—For glass-bottomed or thinner plastic-bottomed (less than<br />

400µ) plates, <strong>iCys</strong> focuses directly on the bottom then moves up by the<br />

bottom thickness plus other offsets to bring the sample into focus.<br />

✧ Default—If you want to allow your users to select a focus method (as<br />

discussed in “Focus and Performance” on page 3-33), select both focus<br />

methods and then select the method that you want to use as a default<br />

setting.<br />

Note: If the Bottom Thickness is less than 400µ, select Bottom Only as the only<br />

available focus method. <strong>iCys</strong> will set the default focus method to match.<br />

❍ Test Autofocus—To test whether you have the values for your carrier set<br />

correctly, you can run Test Autofocus. To test under working conditions,<br />

CompuCyte suggests that you fill the test wells with water or another buffered<br />

solution (such as phosphate-buffered saline [PBS].)<br />

C-20 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B