May 6, 2005 <strong>iCys</strong> Research Imaging Cytometer Hardware Reference<br />

❑ Turning on the brightfield source and viewing the specimen area directly through<br />

the eyepiece.<br />

❑ Using the CCD camera with either the stand alone monitor or the computer<br />

monitor.<br />

➣ In either case, to manually focus the microscope:<br />

1. Lower the objective lens turret to the full downward position.<br />

2. In the Scan Area Setup window, described in “Setting up a Microplate Scan” on<br />

page 3-17 and “Setting up a Freeform Carrier Scan” on page 3-22, select Move<br />

stage to active well ( ), then click on the region of interest to move the stage<br />

into position.<br />

3. Select Instrument ⇒ Brightfield, and switch the microscope Light Path Selector<br />

lever to Observation ( ).<br />

4. If necessary, adjust the brightfield lamp power.<br />

5. From the microscope, rotate the coarse or fine focus knob until the specimen<br />

appears to be in focus on the eyepieces, the camera monitor or the computer<br />

monitor<br />

6. Adjust the focus by the system offset for the objective lens used in the run.<br />

7. The system is now ready to scan the well or scan area.<br />

Objective Lens Automated Movement<br />

When scanning, the microscope automatically moves the objective lens into position<br />

for scanning. When not scanning, there are three possible locations for the objective<br />

lens:<br />

❑ Objective Home—The home base for the lens. The lens is entirely retracted. The<br />

objective moves to the home position when the objective lens is explicitly<br />

returned home. After the objective lens is homed, it is driven to the Clear Position.<br />

❑ Clear position—The lens is retracted enough to be clear of any carriers, but is not<br />

retracted to the Home position. The objective moves to Clear when a Scan & Save<br />

run is completed, is manually stopped, or when the stage is moved to the Load<br />

position.<br />

❑ Retraction offset—A carrier-specific retraction that allows the stage to move<br />

without bumping into the objective. The objective moves to the retraction offset<br />

when:<br />

❍ A Test Scan completes or is stopped.<br />

❍ The stage is moved using the Move to Active Well command between wells<br />

or rooms in the Scan Area Setup window.<br />

❍ You view events in the Field Image window as described in “Locating and<br />

Viewing Events in the Field Image Window” on page 4-45.<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> 2-9