May 6, 2005 <strong>iCys</strong> Research Imaging Cytometer Hardware Reference<br />

Figure 2-10: Light Path Selector Knob<br />

6. Focus the sample using the microscope focus knob or, if available, the rocker<br />

switch and the fine focus knob on the Remote Control unit.<br />

7. After the hardware setup is complete, you can use the <strong>iCys</strong> application to view the<br />

samples on the microscope stage.<br />

Observing Samples using EpiFluorescence<br />

➣ If your <strong>iCys</strong> system includes the optional EpiFluorescence feature, you can use it to<br />

observe samples as follows:<br />

1. Set the light path selector lever to observation ( ).<br />

2. If necessary, deselect Brightfield from the <strong>iCys</strong> Instrument menu.<br />

3. Turn on the Power Supply Unit for the mercury burner.<br />

4. Set the EpiFluorescence cassette to the desired filter position. The following<br />

filters are supplied with <strong>iCys</strong>:<br />

Table 2-1: EpiFluorescence Filters<br />

Turret<br />

Position<br />

Filter Name Filter Short<br />

Name<br />

Eyepieces only<br />

Eyepieces and Camera<br />

(80% binoculars, 20% Camera)<br />

Excitation<br />

Light Color<br />

5. Make sure the shutter on the EpiFluorescence cassette is open:<br />

Camera only<br />

2 U-MNU2 NU Violet Blue<br />

3 UMWIBA2 WIBA Blue Green<br />

4 U-MWIG2 WIG Green Red<br />

Shutter in Open<br />

Position<br />

Emission Color<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> 2-15