<strong>iCys</strong> Cytometric Analysis Software Application Overview <strong>iCys</strong> Task Bar<br />

The following table summarizes the options in the feature field:<br />

Table 3-4: Feature Options<br />

Parameters Meaning<br />

Area Area in square microns of each event.<br />

Background Background fluorescence or signal nearby each event.<br />

Circularity How close the event is to a perfect circle. The circularity is<br />

defined by: perimeter2 /area<br />

Therefore, the circularity of a perfect circle is 4π (12.566).<br />

The larger the circularity number, the less circular it is.<br />

Note that, due to normalization, the circularity of an event<br />

may be less than 12.566.<br />

Id The Id number of the event. Each event is individually tagged<br />

with an Id number.<br />

Integral Integrated fluorescence per event for the selected channel.<br />

MaxPixel Brightest pixel value per event for the selected channel.<br />

Perimeter The perimeter of the event in microns.<br />

Scan Pixel number, from 0 to 768, indicating the y-position of the<br />

scanning mirror where the event was detected.<br />

Time Time, in seconds, at which the event was acquired.<br />

When rescanning a Raw file, the Time parameter contains the<br />

Well Number of this scan area.<br />

WellNo Well number (or scan area number) of event.<br />

X position X position in the carrier for each event.<br />

Y position Y position in the carrier for each event.<br />

3-64 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B