May 6, 2005 <strong>iCys</strong> Cytometric Analysis Software Application Overview<br />

Selecting the Lasers and Channels<br />

About the <strong>iCys</strong> Lasers<br />

<strong>iCys</strong> comes with equipped with up to three lasers:<br />

✧ Argon Ion, which emits at 488 nanometers.<br />

✧ Helium-Neon (HeNe), which emits at 633nm.<br />

✧ Violet, which emits at 405nm.<br />

The lasers you select for a given scan will depend on the excitation characteristics<br />

of the fluorophores used.<br />

Note: Although you can select any laser or laser combination for either the first<br />

or second pass, some laser combinations will not be compatible with the<br />

fluorescence emission wavelengths of the dyes you are using. The<br />

excitation wavelengths should not overlap the emission wavelengths in<br />

which you are interested for any one pass.<br />

Selecting Data Acquisition Channels<br />

❍ Edit—When the Edit button on the Channel tab is selected, you can specify<br />

the lasers and channels to be used in both the first and second passes of the<br />

scan by clicking the Edit button. The following menu will appear.<br />

Revision B <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> 3-41