Generating Analytical Data Locating and Viewing Events in the Field Image Window<br />

If your system does not have AutoFocus, you will need to adjust the focus<br />

by the appropriate offsets for the current action (Scan and/or<br />

Observation).<br />

Locating and Viewing Events using Saved Data<br />

When you are locating events in runs that have already been saved, you can use the<br />

previously saved data.<br />

1. Open a run, selecting “Analyze Events.”<br />

2. When the run is open, use Select Active Well in the Carrier window, as described<br />

in Table 3-1, “Microplate Scan Area Setup Window,” on page 3-18 or Table 3-2,<br />

“Freeform Scan Area Setup Window,” on page 3-24, to determine which set(s) of<br />

data you want to activate.<br />

3. Right-click on the region of interest and select View Cells in Region to open the<br />

View Events window and select the event you want to view, as described in<br />

“Viewing Cell Details” on page 4-35.<br />

4. You can choose to view the selected event in the Field Image window, directly<br />

with the eyepieces, with the camera, or in any combination of the three.<br />

❍ To view the event in the Field Image window, select Scan from the View<br />

Mode box in the View Events window.<br />

❍ To view the event using eyepieces, camera, or both select Observation.<br />

❍ To view the event in the Field Image window and with the eyepieces, camera,<br />

or both, select both Scan and Observation.<br />

5. Click on the row of the event of interest (or default to the first event), then click<br />

View to acquire the image of the highlighted event. <strong>iCys</strong> locates the image as<br />

follows:<br />

❍ If Field Image is selected, the .jpg image of field that contains the event of<br />

interest is displayed in the Field Image window. The live sample is not used<br />

(no stage or microscope objective movement required).<br />

❍ If both Scan and Observation are selected:<br />

✧ The stage moves so the event of interests is in the middle of the field of<br />

view through the eyepieces or color camera monitor.<br />

✧ If your system has AutoFocus, <strong>iCys</strong> moves the microscope objective to<br />

the focus position used during the original scan. If your system does not<br />

have AutoFocus, you will need to manually focus the objective on the<br />

event of interest.<br />

✧ To view the event through the eyepieces or camera monitor, turn the Light<br />

Path Selector Lever to the Observation position ( ).<br />

❍ If Observation is selected:<br />

4-48 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B