Administrator Tasks Defining Custom Carriers<br />

b. Turn the oscilloscope off by clicking the “Play or Stop” button ( ).<br />

c. Press Set from X,Y for the Top Right Room Corner measurement.<br />

d. Turn the oscilloscope on again, open shutter, and adjust the stage until the<br />

bottom (front) left corner of the room is over the objective.<br />

e. Press Set from X,Y for the Bottom Left Room Corner.<br />

If the room is a Circle:<br />

a. Use the touchpad buttons and Remote Control unit X and Y knobs to drive the<br />

stage to a position such that the room is centered over the microscope beam.<br />

b. Turn the oscilloscope off by clicking the “Play or Stop” button ( ).<br />

c. Press Set from X,Y for the Center measurement.<br />

d. Turn the oscilloscope on again and open shutter.<br />

e. For the Diameter measurement, you can measure the room diameter<br />

manually, look up the room size on the specification sheet for this carrier and<br />

enter it for the Diameter value, or use the Remote Control unit X and Y knobs<br />

to drive to the edge of the room and click on Set from X, Y for the Diameter<br />

field.<br />

8. Repeat Step 8 for each room you need to define.<br />

Determining the Freeform Z Parameters<br />

Finding the Base to Well Height<br />

➣ To determine the Base to Well Height:<br />

1. With the stage at room R1, the same objective in place as displayed in the task bar,<br />

and the Oscilloscope window open and on, and the Argon Laser Shutter open, use<br />

the rocker switch on the Remote Control unit to drive the objective upwards<br />

slowly until you observe a reasonably well focused line or lines on the autofocus<br />

monitor. Make sure that you are focusing on the first line or pair of lines to be<br />

imaged on the display.<br />

2. Use the Fine Focus knob on the Remote Control unit to create a well focused<br />

image.<br />

1. Click Set from Z for the Base To Well Height field in the Z tab. Mark down this<br />

value, and then take the same measurement at least once each room on the carrier<br />

(so that you have measurements for a number of positions on the carrier.)<br />

2. Enter the lowest value measured as the Base to Well Height.<br />

C-22 <strong>iCys</strong> Research Imaging Cytometer <strong>User</strong> <strong>Guide</strong> Revision B