<strong>2704</strong> Controller Transducer Scaling<br />

Set the strain gauge bridge to its ‘zeroed’ condition<br />

8. Press as many times<br />

as necessary to scroll to<br />

Scale Low<br />

This will normally be zero<br />

9. Press or to<br />

enter the low end<br />

calibration value<br />

10. Press to scroll to<br />

Scale High<br />

11. Press or to<br />

enter the high end<br />

calibration value<br />

In this example a value of<br />

8000 is chosen which<br />

may represent 80% of the<br />

0 -10,000psi range of a<br />

pressure transducer.<br />

12. Press to scroll to<br />

Start Pnt 1 Cal<br />

13. Press or to<br />

enter On<br />

This parameter can be<br />

configured to be initiated<br />

from a digital input <strong>and</strong><br />

wired, for example, to an<br />

external switch. If this<br />

has been configured this<br />

parameter can be turned<br />

on from this source.<br />

- Tip: To backscroll hold down <strong>and</strong> press<br />

The <strong>controller</strong> automatically performs the procedure described in Section 19-2. During this<br />

time the Cal Active parameter will change to On. When this parameter value changes back to<br />

Off the calibration is complete.<br />

The Shunt State parameter will also change during the procedure to show when it is being<br />

connected (On = connected, Off = disconnected).<br />

Note:-<br />

It is possible to start the calibration procedure before the system has settled at a stable value.<br />

The <strong>controller</strong> continuously takes blocks of 50 samples. When the average value between<br />

two consecutive blocks is within the ‘Threshold Value’ the <strong>controller</strong> will then calibrate.<br />

The Threshold Value defaults to 0.5 but can be adjusted in configuration level. If the<br />

readings are not stable within this period the <strong>controller</strong> will abort the calibration.<br />

Installation <strong>and</strong> Operation H<strong>and</strong>book. Part No HA026502 Issue 1.0 Apr-00 19-5