Transducer Scaling<br />

<strong>2704</strong> Controller<br />

19.6.1. Parameter Notes<br />

1. Enable Cal This may be wired to a digital input for an external switch. If not<br />

wired, then the value may be changed.<br />

When enabled the transducer parameters may be altered as described<br />

in the previous sections. When the parameter has been turned On it<br />

will remain on until turned off manually even if the <strong>controller</strong> is<br />

powered cycled.<br />

2. Start Tare This may be wired to a digital input for an external switch. If not<br />

wired, then the value may be changed.<br />

3. Start Pnt1 Cal This may be wired to a digital input for an external switch. If not<br />

wired, then the value may be changed.<br />

It starts the calibration procedure for:<br />

1. Shunt Calibration<br />

2. The low point for Load Cell Calibration<br />

3. The low point for Comparison Calibration<br />

4. Start Pnt2 Cal This may be wired to a digital input for an external switch. If not<br />

wired, then the value may be changed.<br />

It starts the calibration procedure for:<br />

1. The high point for Load Cell Calibration<br />

2. The high point for Comparison Calibration<br />

5. Clear Cal This may be wired to a digital input for an external switch. If not<br />

wired, then the value may be changed.<br />

When enabled the input will reset to default values. A new calibration<br />

will overwrite the previous calibration values if Clear Cal is not<br />

enabled between calibrations.<br />

6. Threshold Val The input needs to settle within a range which has been set in<br />

configuration level. The threshold value sets the required settling time<br />

for shunt, load cell <strong>and</strong> auto-tare calibration.<br />

7. Shunt This parameter is an output from the function block which can be<br />

wired to a transducer scale module to close the shunt circuit <strong>and</strong><br />

introduce the calibration resistor. It may be used in copy <strong>and</strong> paste<br />

wiring.<br />

19-14 Installation <strong>and</strong> Operation H<strong>and</strong>book. Part No HA026502 Issue 1.0 Apr -00