Transducer Scaling<br />

<strong>2704</strong> Controller<br />

Allow the Process to settle at the high calibration point<br />

13. Press to ‘Start Pnt2<br />

Cal’<br />

14. Press or to ‘On’<br />

This parameter can be<br />

configured to be initiated<br />

from a digital input <strong>and</strong><br />

wired, for example, to an<br />

external switch. If this<br />

has been configured this<br />

parameter can be turned<br />

on from this source.<br />

15. Press as many times<br />

as necessary to scroll to<br />

‘Adjust Value’<br />

16. Press or to<br />

enter the value indicated<br />

on the reference<br />

instrument<br />

17. Press to confirm or<br />

to cancel as<br />

instructed<br />

The confirm message<br />

does not appear unless<br />

‘Adjust Value’ is changed.<br />

If the displayed value is<br />

acceptable change it<br />

momentarily then back to<br />

the value to step to the<br />

next stage.<br />

On confirm the current<br />

input value is stored as<br />

‘Input High’ <strong>and</strong> the value<br />

entered by the user is<br />

stored in the ‘Scale High’<br />

parameter.<br />

It is possible to perform either low or high points in isolation, or to calibrate both points<br />

consecutively as described above.<br />

19-10 Installation <strong>and</strong> Operation H<strong>and</strong>book. Part No HA026502 Issue 1.0 Apr -00