When the Spontaneous Breathing Trial (SBT) mode of ventilation is on and the Total Breath Rate ( f )<br />

is greater than the SBT HIGH f setting (high limit of breath rate range) for 30 seconds, the SBT > f<br />

alarm is generated.<br />

For additional information concerning the Spontaneous Breathing Trial (SBT) mode of ventilation, see<br />

SBT (Spontaneous Breathing Trial) Operations in Chapter 10 - Extended Features.<br />

When a SBT > f alarm is generated:<br />

The SBT > f message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the SBT > f alarm:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm (silences the audible alarm and clears<br />

the flashing display).<br />

When the SBT > f alarm is reset (Silence Reset button pushed twice), the alarm is<br />

suspended for the next 60 seconds.<br />

The 60-second suspension of the SBT > f alarm is only enabled when the alarm is manually<br />

silenced/reset by pushing the Silence Reset button twice. It is not enabled when the SBT > f<br />

alarm is automatically silenced/reset because the patient’s breath rate is no longer greater<br />

than the set SBT HIGH f alarm value.<br />

When an SBT alarm has been active in excess of 5 minutes, the SBT mode of ventilation is<br />

terminated, the ventilator clears the alarm status, silences the audible alarm and returns to its<br />

previous mode of ventilation / settings. To remove the flashing SBT alarm message from the<br />

display window, push the Silence Reset button.<br />

NOTE<br />

The Spontaneous Breathing Trial feature is only available when Flow Sensing Patient Circuit mode is<br />

enabled (see Chapter 12 – Operating Procedure for additional information).<br />

REMARQUE<br />

La fonction d’essai de respiration spontanée n'est disponible que lorsque le mode de circuit patient<br />

de détection d'écoulement est activé (reportez-vous au chapitre 12 – Procédure d'utilisation pour<br />

d'autre information).<br />