If the Patient Circuit Mode Query is enabled/on when the ventilator is powered up, ventilation and<br />

alarm activation are suspended and the message FLOW SENSOR or NO FL SENSOR is displayed<br />

(see Queries in Chapter 10 - Extended Features for additional information).<br />

<br />

To enable the suspended alarms and begin ventilation with the settings in use during<br />

the last power cycle, turn the Set Value knob to the desired patient circuit mode<br />

(FLOW SENSOR or NO FL SENSOR) and press the Select button (see Patient<br />

Circuit Mode in Chapter 12 – Operating Procedure for additional information).<br />

If no controls are activated for three (3) seconds while either the Leak Test Query or Patient Circuit<br />

Mode Query options are being displayed, an audible alert sounds. Activation of any control resets the<br />

3 second delay of the audible alert.<br />

To prevent autocycling, the Leak Compensation feature (if enabled/on) is suspended during the first<br />

30 seconds of operation.<br />

To prevent nuisance alarms, the LOW MIN VOL alarm (Low Minute Volume) is suspended for the<br />

first 20 seconds and the HIGH f alarm (High Breath Rate) is suspended for the first 60 seconds of<br />

operation.<br />

If the Power On Self Tests fail;<br />

The mode of failure (CPU, SRAM, INT VECTOR, ROM CRC or EEPROM) is displayed in the<br />

message window and an audible alarm sounds continuously.<br />

Turn the ventilator off by pushing the On/Standby button<br />

Silence the alarm by pushing the Silence Reset button<br />

Discontinue use of the ventilator and immediately contact a certified <strong>CareFusion</strong> service<br />

technician or <strong>CareFusion</strong><br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 12-5