06.02.2014 Views

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Sigh Breath<br />

Use the Sigh Breath feature to deliver a sigh breath every 99 machine breaths or every 7 minutes,<br />

whichever comes first. A sigh breath is delivered at 1.5 times the set Tidal Volume and an inspiration<br />

time of 1.5 times the set Inspiratory Time.<br />

The High Pressure Alarm limit is increased 1.5 times (100 cmH 2 O maximum) when a sigh breath<br />

occurs.<br />

When the Sigh Breath feature is enabled/on, the message SIGH ON is displayed in the display<br />

window in addition to the normal monitored values.<br />

To modify the SIGH setting:<br />

5) Push the Select button while SIGH is displayed.<br />

6) SIGH ON or SIGH OFF is displayed.<br />

7) Turn the Set Value knob until the desired state is displayed.<br />

8) Push the Select button.<br />

Options: ON or OFF<br />

10-20 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E