To Remove the Lower Protective Boot 76 :<br />

1) Lay the ventilator down (front up) and use a Phillips-head screwdriver to remove the four flathead<br />

mounting screws and finish washers in the sides of the lower protective boot (2), as<br />

indicated in the illustration.<br />

2) Remove the lower boot (2) and insert and thread four #4-40 X 1/4” flat-head mounting screws (5)<br />

with finish washers (6) into the screw holes in the ventilator’s side panels, as indicated in the<br />

illustration.<br />

Finish washers (6) should be already in place.<br />

3) Torque tighten all four screws to 20 in-oz (0.14 Nm) (do not over tighten to avoid damage to the<br />

finish washers).<br />

Item (1), Protective Boot, Upper (1), P/N 19033-001 (Purple)<br />

Item (2), Protective Boot, Lower (1), P/N 19032-001 (Purple)<br />

Item (3), #4-40 X 3/16” Pan-head screw (1), P/N 10438 77<br />

Item (4), #4-40 X 1/4” Pan-head mounting screws (2), P/N 10435 77<br />

Item (5), #4-40 X 1/4” Flat-head mounting screws (6), P/N 10430 77<br />

Item (6), Finish Washers (6), P/N 19119-002 77<br />

76 Refer to Page C-9 for information concerning the appropriate accessories mounting screws or accessories<br />

replacement screws location, type and length to use when removing or exchanging external accessories on an<br />

<strong>LTV</strong> ® ventilator.<br />

77 Contained in <strong>CareFusion</strong> Replacement Screws kit, P/N 11149.<br />

21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator C-5