REMOVE PTNT<br />

When the ventilator is powered up in the Ventilator Checkout or Ventilator Maintenance modes, the<br />

REMOVE PTNT alarm is generated to remind you to remove the patient from the ventilator before<br />

proceeding. Use the Ventilator Checkout mode to check for correct operation of the displays and<br />

controls and to check the patient circuit for leaks. Ventilator Maintenance mode is used by technical<br />

personnel to perform maintenance or calibration.<br />

WARNING<br />

Ventilator Checkout and Maintenance Modes - The <strong>LTV</strong> ® <strong>1100</strong> ventilator does not deliver gas<br />

during the Ventilator Checkout mode (VENT CHECK) or Ventilator Maintenance mode (VENT<br />

MTNCE) and should not be used to ventilate a patient during these tests.<br />

AVERTISSEMENT<br />

Modes Vérification et Entretien du ventilateur - Le ventilateur de la <strong>1100</strong> <strong>LTV</strong> ® ne transmet pas le<br />

mélange de gaz en mode Vérification du ventilateur (VENT CHECK) ou en mode Entretien du<br />

ventilateur (VENT MTNCE), il ne devrait donc pas être utilisé pour ventiler un patient durant<br />

l'exécution de ces tests.<br />

When you enter Ventilator Checkout mode or Ventilator Maintenance mode, a REMOVE PTNT<br />

alarm occurs:<br />

The REMOVE PTNT message is displayed.<br />

The audible alarm is sounded.<br />

To reset the REMOVE PTNT alarm:<br />

1) Push the Silence Reset button twice.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 9-23