SBT TIME<br />

The SBT TIME message is displayed during the Spontaneous Breathing Trial mode of ventilation<br />

when only two (2) minutes of SBT mode of ventilation remain.<br />

This is an informational message only.<br />

When a SBT TIME message is generated:<br />

The SBT TIME message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the SBT TIME message:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm status message (silences the audible<br />

alarm and clears the flashing display).<br />

WARMUP xx<br />

When the ventilator is first powered up, the transducers require up to 60 seconds of warm-up time<br />

before they will operate within their normal tolerances. During this warm-up period, the ventilator will<br />

not allow you to run the leak test or calibration. If you select an option that is not available during the<br />

warm-up period, the WARMUP xx message is displayed. When the warm-up period has expired, the<br />

message is removed.<br />

When a WARMUP message occurs:<br />

The WARMUP message and the remaining warm-up time are displayed in the window.<br />

The ventilator does not allow the restricted functions to be performed.<br />

To reset the WARMUP message:<br />

1) The WARMUP message will automatically reset when the warm-up period has expired.<br />

9-36 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E