High PRES (cont.)<br />

The HIGH PRES alarm becomes inactive and is automatically silenced using the following criteria:<br />

High Pressure<br />

Limit Setting<br />

Circuit Pressure at which HIGH PRES alarm is silenced<br />

31 to 100 cmH 2 O Less than 25 cmH 2 O<br />

8 to 30 cmH 2 O More than 5 cmH 2 O lower than current High Pressure Limit Setting<br />

5 to 7 cmH 2 O Less than 2 cmH 2 O<br />

When a HIGH PRES alarm occurs:<br />

Inspiration is immediately terminated and exhalation begins.<br />

The HIGH PRES message is flashed in the display window.<br />

The High Pressure Limit control display is flashed.<br />

The audible alarm is sounded.<br />

To reset the HIGH PRES alarm:<br />

1) Push the Silence Reset button to silence the alarm.<br />

2) Resolve the high pressure problem.<br />

3) Push the Silence Reset button to reset the alarm.<br />

9-14 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E