SBT OFF<br />

When the minutes preset in the SBT OP, MINUTES menu have elapsed and a Spontaneous<br />

Breathing Trial (SBT) mode of ventilation is terminated, the SBT OFF alarm is generated.<br />

For additional information concerning the Spontaneous Breathing Trial (SBT) mode of ventilation, see<br />

SBT (Spontaneous Breathing Trial) Operations in Chapter 10 - Extended Features.<br />

When a SBT OFF alarm is generated:<br />

The SBT OFF message is flashed in the display window.<br />

The audible alarm is sounded for 3 seconds.<br />

To clear the SBT OFF flashing display:<br />

1) Push the Silence Reset button.<br />

NOTE<br />

The Spontaneous Breathing Trial feature is only available when Flow Sensing Patient Circuit mode is<br />

enabled (see Chapter 12 – Operating Procedure for additional information).<br />

REMARQUE<br />

La fonction d’essai de respiration spontanée n'est disponible que lorsque le mode de circuit patient<br />

de détection d'écoulement est activé (reportez-vous au chapitre 12 – Procédure d'utilisation pour<br />

d'autre information).<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 9-29