Procedure:<br />

Set the Low Min. Vol. alarm to 10 L C-19<br />

2) B)<br />

Reset the Low Min. Vol. alarm to<br />

1.0 and clear the alarm.<br />

Set the Low Pressure alarm to 60.<br />

Set the Low Pressure alarm to 5<br />

and clear the alarm.<br />

Set the High Pres Limit to 10<br />

cmH 2 O below the Peak Inspiratory<br />

Pressure.<br />

Reset the High Pres. Limit alarm to<br />

100 and clear the alarm.<br />

Disconnect the high pressure sense<br />

line from the ventilator<br />

Reconnect the high pressure sense<br />

line and clear the alarm<br />

Disconnect AC Adapter from<br />

Ventilator at pigtail cable connector.<br />

LOW MIN VOL alarm activates<br />

C-19<br />

2) C) LOW PRES alarm activates<br />

C-19<br />

2) D)<br />

C-19<br />

2) E)<br />

C-19<br />

2) F)<br />

C-20<br />

2) G)<br />

C-20<br />

2) H)<br />

HIGH PRES alarm activates.<br />

DISC/SENSE alarm activates on the next<br />

breath<br />

POWER LOST alarm activates<br />

Battery Level LED illuminates showing the<br />

charge level.<br />

Ventilator continues to operate from the<br />

internal battery<br />

C-22 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E