Volume Control Breaths<br />

For Volume Control breaths, the set Tidal Volume is delivered over the set Inspiratory Time and flow<br />

is delivered in a decelerating taper flow waveform. Peak flow is calculated based on the Tidal Volume<br />

and Inspiratory Time and the final flow is 50% of the peak flow. Volume breaths may be machine or<br />

assist type breaths.<br />

When the combination of inspiratory time and tidal volume result in an initial flow of