High f<br />

Use this menu item to set the high breath rate and time period alarm values. When the Total Breath<br />

Rate ( f ) exceeds the set high breath rate and time period alarm values, an audible alarm will be<br />

sounded and a flashing HIGH f message will be displayed.<br />

To set the high breath rate and time period alarm values:<br />

1) Push the Select button while HIGH f is displayed and f is displayed.<br />

2) Push the Select button while f is displayed and HIGH f OFF or f xx bpm is displayed.<br />

3) Turn the Set Value knob until the desired setting is displayed, push the Select button and the<br />

high breath rate alarm value is set.<br />

Range: 5 - 80 bpm (in increments of 1) - HIGH f OFF<br />

4) Turn the Set Value knob until TIME is displayed, push the Select button and xx sec is displayed.<br />

5) Turn the Set Value knob until the desired setting is displayed and push the Select button. The<br />

high breath rate time period alarm value is set.<br />

Range: 0 - 60 seconds, in increments of 10<br />

High PEEP<br />

Use this menu item to set a high PEEP alarm. When the current PEEP value exceeds the set high<br />

PEEP alarm value, an audible alarm will be sounded and a flashing HIGH PEEP message will be<br />

displayed.<br />

To modify the HIGH PEEP alarm value:<br />

1) Push the Select button while HIGH PEEP is displayed.<br />

2) HI PEEP OFF or PEEP +XX cmH 2 0 is displayed.<br />

3) Turn the Set Value knob until the desired setting is displayed.<br />

4) Push the Select button.<br />

Range: PEEP +3 through PEEP +20 cmH 2 0, HI PEEP OFF<br />

Low PEEP<br />

Use this menu item to set a low PEEP alarm. When the current PEEP value is less than the set low<br />

PEEP alarm value, an audible alarm will be sounded and a flashing LOW PEEP message will be<br />

displayed.<br />

To modify the LOW PEEP alarm value:<br />

1) Push the Select button while LOW PEEP is displayed.<br />

2) LO PEEP OFF or PEEP -XX cmH 2 0 is displayed.<br />

3) Turn the Set Value knob until the desired setting is displayed.<br />

4) Push the Select button.<br />

Range: PEEP -3 through PEEP -20 cmH 2 0, LO PEEP OFF<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 10-5