Symptoms Possible Causes What to Do<br />
continued…<br />
Repeated<br />
DISC/SENSE alarms.<br />
Repeated XDCR<br />
FAULT alarms.<br />
HW FAULT alarm<br />
RESET alarm occurs<br />
after ventilator is<br />
operated on internal<br />
battery until it is fully<br />
depleted.<br />
RESET, CRC,<br />
STACK, POST, or<br />
RUNAWAY alarms<br />
NO CAL DATA alarm.<br />
NO CAL displayed in<br />
place of monitored<br />
values.<br />
Exhalation drive line leaking<br />
or loose.<br />
Exhalation valve leaking<br />
during inspiration.<br />
Internal problem with the<br />
ventilator.<br />
Internal problem with the<br />
ventilator.<br />
Electro static discharge<br />
(ESD).<br />
Fan was bumped or<br />
temporarily stopped while<br />
cleaning fan filter.<br />
Internal problem with the<br />
ventilator.<br />
Internal battery depleted.<br />
Electro static discharge<br />
(ESD).<br />
Internal problem with the<br />
ventilator.<br />
Failed or missing calibration<br />
records.<br />
Check the exhalation drive line at both<br />
the ventilator and exhalation valve ends.<br />
Verify the line is securely seated and not<br />
leaking.<br />
Verify the exhalation valve is not leaking<br />
during inspiration. If it is leaking, open<br />
the exhalation valve and remove the<br />
diaphragm. Reseat the diaphragm. See<br />
page 13-8 for a diagram of correct<br />
exhalation valve assembly.<br />
If necessary, replace the exhalation<br />
diaphragm, or exhalation valve with a<br />
new one.<br />
Immediately contact a certified<br />
<strong>CareFusion</strong> service technician.<br />
Immediately contact a certified<br />
<strong>CareFusion</strong> service technician.<br />
Clear the alarm. Reduce static causing<br />
conditions in the operating environment.<br />
Clear the alarm. No further action<br />
required if alarm does not reoccur.<br />
If problem reoccurs, immediately contact<br />
a certified <strong>CareFusion</strong> service<br />
technician.<br />
This is a normal. Clear the alarm and<br />
charge the internal battery (see page 14-<br />
10 for instructions on charging the<br />
internal battery).<br />
Clear the alarm. Reduce static causing<br />
conditions in the operating environment.<br />
If problem reoccurs, immediately contact<br />
a certified <strong>CareFusion</strong> service<br />
technician.<br />
Immediately contact a certified<br />
<strong>CareFusion</strong> service technician.<br />
15-20 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E