Charge Status<br />

The Charge Status indicator shows the charge state of the internal battery. This LED is on when the<br />

ventilator is supplied with external power and the internal battery is being charged. The charge status<br />

is indicated as follows:<br />

LED Color<br />

Flashing Amber<br />

Green<br />

Amber<br />

Red<br />

Charge Status<br />

The ventilator is performing pre-charge testing of the battery before<br />

starting the charge process. This occurs when external power is first<br />

connected. The process normally takes a few seconds but may take up to<br />

an hour on a deeply discharged battery.<br />

The internal battery is charged to full level. While in this state, the charger<br />

will continue to trickle charge the battery.<br />

The internal battery is being bulk charged. The battery has not reached a<br />

full charge level yet.<br />

The ventilator has detected a charge internal battery fault. The internal<br />

battery cannot be charged.<br />

CAUTION<br />

Charge Fault - If the Charge Status LED indicates a charge fault, contact a certified <strong>CareFusion</strong><br />

service technician immediately.<br />

Internal Battery Use: The internal battery is intended for use during short periods while switching<br />

between external power supplies, in emergencies or for short duration transports. The length of time<br />

the ventilator will operate on internal power is a function of factors such as settings, charge level and<br />

condition of the battery; therefore, the use of the internal battery as a standard operating practice is<br />

not recommended.<br />

ATTENTION<br />

Erreur de charge - Si le DEL de l'état de charge indique une erreur de charge, veuillez contacter<br />

immédiatement un technicien de service certifié <strong>CareFusion</strong>.<br />

Utilisation de la batterie interne: La batterie interne est conçue pour être utilisée sur de courtes<br />

périodes pendant la commutation entre des connexions d’alimentation externe, les situations<br />

d’urgence ou les transports de courte durée. La durée pendant laquelle le ventilateur fonctionnera sur<br />

l’alimentation interne dépend de facteurs tels, la configuration, le niveau de la charge et la condition<br />

de la batterie; l’utilisation de la batterie interne pour l'opération normale n’est donc pas<br />

recommandée.<br />