Variable Rise Time<br />

Use the Variable Rise Time option to select the rise time profile for Pressure Support breaths. The<br />

rise time profiles are numbered 1 through 9 where 1 is the fastest rise time and 9 is the slowest rise<br />

time. Starting with the fastest rise time, each time is 33% longer than the previous one.<br />

To modify the Rise Time Profile:<br />

1) Push the Select button while RISE TIME is displayed.<br />

2) PROFILE x is displayed.<br />

3) Turn the Set Value knob until the desired Rise Time Profile is displayed.<br />

4) Push the Select button.<br />

Range: 1 to 9, where 1 is the fastest and 9 is the slowest<br />

10-8 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E