Sensitivity<br />

Use the Sensitivity control to establish the threshold level to allow the patient to flow trigger delivered<br />

breaths.<br />

A flow trigger occurs when:<br />

The O 2 Conserve feature is OFF (CONSERVE OFF),<br />

And The ventilator is operating in Flow Sensing Patient Circuit mode,<br />

And The Sensitivity is set to any value from 1 to 9 (Sensitivity value is in lpm, when O 2 Conserve<br />

is off and the ventilator is operating in Flow Sensing Patient Circuit mode),<br />

And the ventilator is in exhalation phase,<br />

And the minimum exhalation time has expired,<br />

And the flow is greater than or equal to the Sensitivity setting<br />

The LEAK measurement displayed in the RT XDCR DATA menu can be used to help select an<br />

appropriate sensitivity value. Typically, the sensitivity value is set higher than the displayed LEAK<br />

measurement. For instance, if the LEAK measurement were up to 2.53, a minimum sensitivity of<br />

three (3) would be appropriate.<br />

A pressure trigger occurs when:<br />

The O 2 Conserve feature is ON (CONSERVE ON) and/or the ventilator is operating in<br />

Non-Flow Sensing Patient Circuit mode,<br />

And The Sensitivity is set to any value from 1 to 9 (Sensitivity value is in cmH 2 O, when O 2<br />

Conserve is on and/or the ventilator is operating in Non-Flow Sensing Patient Circuit mode),<br />

And the ventilator is in exhalation phase,<br />

And the minimum exhalation time has expired,<br />

And the airway pressure drops below or equal to the Sensitivity setting<br />

Backup pressure triggers are enabled when the setting is any value other than a dash “-”.<br />

A backup pressure trigger occurs when:<br />

The Sensitivity is set to any value from 1 to 9,<br />

And the ventilator is in exhalation phase,<br />

And the minimum exhalation time has expired,<br />

And the airway pressure drops below -3 cmH 2 O<br />

When a trigger is detected, the Patient Effort LED is illuminated briefly.<br />

NOTE<br />

Triggers are disabled when the Sensitivity setting is set to “-”.<br />

REMARQUE<br />

Les amorces sont désactivées lorsque le réglage de la sensibilité est réglé sur « - ».<br />

To set Sensitivity:<br />

1) Push the Sensitivity button.<br />

2) Change the setting using the Set Value knob.<br />

Range: 1 - 9, “-”, 1 is the most sensitive, 9 is the least sensitive and “-“ is off.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 6-13