Sensitivity<br />
Use the Sensitivity control to establish the threshold level to allow the patient to flow trigger delivered<br />
breaths.<br />
A flow trigger occurs when:<br />
The O 2 Conserve feature is OFF (CONSERVE OFF),<br />
And The ventilator is operating in Flow Sensing Patient Circuit mode,<br />
And The Sensitivity is set to any value from 1 to 9 (Sensitivity value is in lpm, when O 2 Conserve<br />
is off and the ventilator is operating in Flow Sensing Patient Circuit mode),<br />
And the ventilator is in exhalation phase,<br />
And the minimum exhalation time has expired,<br />
And the flow is greater than or equal to the Sensitivity setting<br />
The LEAK measurement displayed in the RT XDCR DATA menu can be used to help select an<br />
appropriate sensitivity value. Typically, the sensitivity value is set higher than the displayed LEAK<br />
measurement. For instance, if the LEAK measurement were up to 2.53, a minimum sensitivity of<br />
three (3) would be appropriate.<br />
A pressure trigger occurs when:<br />
The O 2 Conserve feature is ON (CONSERVE ON) and/or the ventilator is operating in<br />
Non-Flow Sensing Patient Circuit mode,<br />
And The Sensitivity is set to any value from 1 to 9 (Sensitivity value is in cmH 2 O, when O 2<br />
Conserve is on and/or the ventilator is operating in Non-Flow Sensing Patient Circuit mode),<br />
And the ventilator is in exhalation phase,<br />
And the minimum exhalation time has expired,<br />
And the airway pressure drops below or equal to the Sensitivity setting<br />
Backup pressure triggers are enabled when the setting is any value other than a dash “-”.<br />
A backup pressure trigger occurs when:<br />
The Sensitivity is set to any value from 1 to 9,<br />
And the ventilator is in exhalation phase,<br />
And the minimum exhalation time has expired,<br />
And the airway pressure drops below -3 cmH 2 O<br />
When a trigger is detected, the Patient Effort LED is illuminated briefly.<br />
NOTE<br />
Triggers are disabled when the Sensitivity setting is set to “-”.<br />
REMARQUE<br />
Les amorces sont désactivées lorsque le réglage de la sensibilité est réglé sur « - ».<br />
To set Sensitivity:<br />
1) Push the Sensitivity button.<br />
2) Change the setting using the Set Value knob.<br />
Range: 1 - 9, “-”, 1 is the most sensitive, 9 is the least sensitive and “-“ is off.<br />
21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 6-13