LOW MIN VOL<br />

When the exhaled minute volume (VE) is less than the Low Minute Volume setting, the LOW MIN<br />

VOL alarm is generated.<br />

<br />

To prevent nuisance alarms, the LOW MIN VOL alarm is suspended for the first 20 seconds of<br />

ventilator operation after power up and passing the Power On Self Tests.<br />

NOTE<br />

The Low Minute Volume alarm setting is only active and adjustable when Flow Sensing Patient<br />

Circuit mode is enabled. The alarm is inactive and the display window is set to dashes “- -“ when<br />

Non-Flow Sensing Patient mode is enabled (see Chapter 12 – Operating Procedure for additional<br />

information).<br />

REMARQUE<br />

Le réglage de l'alarme de faible débit-volume n'est actif et ajustable que lorsque le mode de circuit<br />

patient de détection d'écoulement est activé. L'alarme est inactive et la fenêtre d'affichage est réglée<br />

à des tirets « - » lorsque le mode de circuit patient de non détection de circulation est activé<br />

(reportez-vous au chapitre 12 - Procédure d'utilisation pour d'autre information).<br />

When a LOW MIN VOL alarm occurs:<br />

The LOW MIN VOL message is flashed in the display window.<br />

The Low Min. Vol. control display is flashed.<br />

The audible alarm is sounded.<br />

To reset the LOW MIN VOL alarm:<br />

1) Push the Silence Reset button twice.<br />

WARNING<br />

Low Minute Volume Control Settings - The Low Min. Vol. control should be set to its highest<br />

clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower<br />

values or off (“- -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse<br />

Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.<br />

AVERTISSEMENT<br />

Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être<br />

ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être<br />

ajustée à des valeurs inférieures ou mise à l’arrêt (“- -“) pour satisfaire aux besoins cliniques,<br />

effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d.,<br />

sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio-respiratoire) s’avère pertinente.<br />