Silence Reset<br />

Use this button to silence an alarm for 60 seconds, to reset an alarm, to start a 60 second preemptive<br />

silence period, and to permanently silence the Vent Inop and Standby alarms. Two important<br />

definitions for understanding how the Silence Reset button works:<br />

Active alarm: An alarm for which the condition currently exists.<br />

Inactive alarm: An alarm that has occurred, but for which the condition no longer<br />

exists.<br />

Silencing and Clearing Alarms:<br />

To silence an active alarm for 60 seconds:<br />

1) Push the Silence Reset button. The audible alarm will be silenced for 60 seconds. Once the<br />

silence period expires, the audible alarm will resume sounding.<br />

To clear an inactive alarm:<br />

1) Push the Silence Reset button. The visual alarm displays will be cleared.<br />

To cancel an active alarm:<br />

1) Push the Silence Reset button twice. The audible alarm will be silenced and the visual alarm<br />

displays will be cleared and the silence period will be terminated.<br />

Preemptive Silence Period<br />

To start a preemptive silence period:<br />

1) Push the Silence Reset button. A 60 second silence period will begin. For any alarms that occur<br />

during the silence period, the visual displays will flash, but the audible alarm will remain silenced<br />

until the end of the silence period.<br />

Vent Inop and Standby alarms:<br />

To silence the Vent Inop or Standby alarm:<br />

1) Push the Silence Reset button. The audible alarm will be permanently silenced, but the Vent<br />

Inop LED will remain lit for a minimum of 5 minutes. This does not adversely affect battery life.<br />